Kendall 1010P and Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes (2014-09-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicableB. Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
40000006
Not applicableA. Kendall 1010P Multi-Function Defibrillation Electrodes
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
22660PC

Introduction

Recalled products
  1. Kendall 1010P Multi-Function Defibrillation Electrodes
  2. Medi-Trace Cadence Adult Multi-Function Defibrillation Electrodes

Reason

Incompatibility between the Philips Automated External Defibrillators (AED) models FR3 and FRX defibrillators and Covidien Medi-Trace cadence and Kendall Defibrillation Electrodes. These electrodes will not connect with Philips FR3 and FRX AED units, and in the case of the use of Covidien Defibrillation Electrodes with the Philips FR3 AED units, could result in a delay of therapy. The FRX AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use.