Multidiagnost Eleva - Flat Detector - Image Detector, Allura XPER FD10/10 - Main Unit,Allura XPER FD20/20-Main Unit,Allura XPER FD10 and Allura XPER FD20 System - Main Unit (2014-09-11)
A) Multidiagnost Eleva - Flat Detector - Image Detector
0708MD02
Introduction
Recalled Products
A) Multidiagnost Eleva - Flat Detector - Image Detector
B) Allura XPER FD10/10 - Main Unit
C) Allura XPER FD20/20-Main Unit
D) Allura XPER FD10
E) Allura XPER FD20 System - Main Unit
Reason
Philips has become aware that if a footswitch is frequently used on an anti-fatigue mat, on a not flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.