A) Advia Centaur System - HBC Total Assay
B) Advia Centaur System XP - HBC Total Assay
C) Advia Centaur System CP - HBC Total Assay
Reason
Siemens Healthcare Diagnostics has confirmed increased assay imprecision and a positive bias (increase in index values) in the low end of the assay with the Advia Centaur Systems HBC total (HBCT) assay, reagent lot 100064. The increased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in false reactive results