Health Products

Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device (2014-09-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
10140000-1146
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device10501331

Introduction

Recalled Products

Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device

Reason

The stereo biopsy devices (serial numbers 2000 through 2090) for Mammomat Inspiration Mammography Systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already inserted, it could lead to an unintentional patient injury.