Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device
10501331
Introduction
Recalled Products
Mammomat Inspiration Mammography System with Stereostatic/Biopsy Device
Reason
The stereo biopsy devices (serial numbers 2000 through 2090) for Mammomat Inspiration Mammography Systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already inserted, it could lead to an unintentional patient injury.