Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes and PA230E Variable Band Phased Array (2014-08-25)
B) Esaote PA230E Transducer used with Esaote MyLab Systems- PA230E Variable Band Phased Array
Manufacturer
Esaote Europe BV,
Philipsweg 1,
Maastricht, Netherlands
6227AJ
9600165000
More than 10 numbers, contact manufacturer.
A) Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes
Manufacturer
Esaote Europe BV,
Philipsweg 1,
Maastricht, Netherlands
6227AJ
9600165000
Introduction
Recalled Products
A) Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes
B) Esaote PA230E Transducer used with Esaote MyLab Systems- PA230E Variable Band Phased Array
Reason
Using the PA230E transducer with a user-defined setting with the output power to 100% (max) in 2D-TEI mode focusing at maximum depth, could cause the transducer surface temperature to exceed the maximum allowable limit.