Health Products

Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes and PA230E Variable Band Phased Array (2014-08-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B) Esaote PA230E Transducer used with Esaote MyLab Systems- PA230E Variable Band Phased Array
Manufacturer
Esaote Europe BV,
Philipsweg 1,
Maastricht, Netherlands
6227AJ
9600165000
More than 10 numbers, contact manufacturer.A) Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes
Manufacturer
Esaote Europe BV,
Philipsweg 1,
Maastricht, Netherlands
6227AJ
9600165000

Introduction

Recalled Products

A) Esaote PA230E Transducer used with Esaote MyLab Systems- MyLab40 Ultrasound System Phased Array Probes
B) Esaote PA230E Transducer used with Esaote MyLab Systems- PA230E Variable Band Phased Array

Reason

Using the PA230E transducer with a user-defined setting with the output power to 100% (max) in 2D-TEI mode focusing at maximum depth, could cause the transducer surface temperature to exceed the maximum allowable limit.