UniCel DxI 600, DxI 800 Access Immunoassay Systems and UniCel DxC 880I, DxC 660I, DxC 680I, DxC 860I Synchron Access Clinical Systems (2014-08-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
600785
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
H) UniCel DxC 860I Synchron Access Clinical Systems4773
900359
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
I) UniCel DxI 600 Access Immunoassay Systems(DL 75053)A71461
900613
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
F) UniCel DxC 660I Synchron Access Clinical Systems4771
900613
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
G) UniCel DxC 680i Synchron Access Clinical Systems4772
600785
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
D) UniCel DxI 800 Access Immunoassay Systems (DL 62733)973100
600785
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
E) UniCel DxI 800 Access Immonuassay Systems (DL 62132)973100
900613
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
K) UniCel DxI 600 Access Immunoassay Systems (DL 89063)A30260
900613
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B) UniCel DxI 600 Access Immunoassay Systems (DL 77901)A30260
600785
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C) UniCel DxC 880I Synchron Access Clinical Systems4768
600785
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
L) UniCel DxI 800 Access Immunoassay Systems(DL 89064)973100
900613
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A) UniCel DxI 600 Access Immunoassay Systems (DL 75053)A30260
900359
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
J) UniCel DxI 600 Access Immunoassay Systems (DL 77901)A71461

Introduction

Recalled Products

A) UniCel DxI 600 Access Immunoassay Systems (DL 75053)
B) UniCel DxI 600 Access Immunoassay Systems (DL 77901)
C) UniCel DxC 880I Synchron Access Clinical Systems
D) UniCel DxI 800 Access Immunoassay Systems (DL 62733)
E) UniCel DxI 800 Access Immonuassay Systems (DL 62132)
F) UniCel DxC 660I Synchron Access Clinical Systems
G) UniCel DxC 680i Synchron Access Clinical Systems
H) UniCel DxC 860I Synchron Access Clinical Systems
I) UniCel DxI 600 Access Immunoassay Systems(DL 75053)
J) UniCel DxI 600 Access Immunoassay Systems (DL 77901)
K) UniCel DxI 600 Access Immunoassay Systems (DL 89063)
L) UniCel DxI 800 Access Immunoassay Systems(DL 89064)

Reason

Beckman coulter has determined that nonconforming sample probe cables may have been installed on newly manufactured and/or recently serviced UniCel DxI instruments. Use of a nonconforming sample probe cable can cause a delay in reporting results, but will not cause erroneous results. A nonconforming sample probe cable may cause level sense errors. If a level sense error occurs due to a nonconforming cable, the test may not start or an in-process test is automatically cancelled.