A) UniCel DxI 600 Access Immunoassay Systems (DL 75053)
A30260
900359
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
J) UniCel DxI 600 Access Immunoassay Systems (DL 77901)
A71461
Introduction
Recalled Products
A) UniCel DxI 600 Access Immunoassay Systems (DL 75053)
B) UniCel DxI 600 Access Immunoassay Systems (DL 77901)
C) UniCel DxC 880I Synchron Access Clinical Systems
D) UniCel DxI 800 Access Immunoassay Systems (DL 62733)
E) UniCel DxI 800 Access Immonuassay Systems (DL 62132)
F) UniCel DxC 660I Synchron Access Clinical Systems
G) UniCel DxC 680i Synchron Access Clinical Systems
H) UniCel DxC 860I Synchron Access Clinical Systems
I) UniCel DxI 600 Access Immunoassay Systems(DL 75053)
J) UniCel DxI 600 Access Immunoassay Systems (DL 77901)
K) UniCel DxI 600 Access Immunoassay Systems (DL 89063)
L) UniCel DxI 800 Access Immunoassay Systems(DL 89064)
Reason
Beckman coulter has determined that nonconforming sample probe cables may have been installed on newly manufactured and/or recently serviced UniCel DxI instruments. Use of a nonconforming sample probe cable can cause a delay in reporting results, but will not cause erroneous results. A nonconforming sample probe cable may cause level sense errors. If a level sense error occurs due to a nonconforming cable, the test may not start or an in-process test is automatically cancelled.