B) Vitros 5600 Integrated System (Device Licence #80701)
680 2413
56000576
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A) Vitros 5600 Integrated System (Device Licence #78571)
680 2413
Introduction
Recalled Products
A) Vitros 5600 Integrated System (Device Licence #78571)
B) Vitros 5600 Integrated System (Device Licence #80701)
Reason
Ortho Clinical Diagnostics (OCD) has identified an anomaly with Vitros system software version 3.1. Internal testing confirmed that when using calibrator barcode labels supplied with Vitros chemistry products calibrator kit 2, calibration may not occur under specific circumstances. Consignees are being advised not to use the calibrator barcode labels for Vitros calibrator kit 2 until this anomaly is resolved in a future version of software.