Health Products

UniCel DxC 600I Synchron Access Clinical Analyzer and Access 2 Immunoassay Systems (2014-09-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.D) Access 2 Immunoassay Systems
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
81600N
More than 10 numbers, contact manufacturer.B) Access 2 Immunoassay Systems- Analyzer Class III
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
81600N
More than 10 numbers, contact manufacturer.C) Access 2 Immunoassay Systems- Analyzer Class II
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
81600N
800074A) UniCel DxC 600I Synchron Access Clinical Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
4767

Introduction

Recalled Products

A) UniCel DxC 600I Synchron Access Clinical Analyzer
B) Access 2 Immunoassay Systems- Analyzer Class III
C) Access 2 Immunoassay Systems- Analyzer Class II
D) Access 2 Immunoassay Systems

Reason

Beckman coulter has determined that the Access 2 console PC (B23083) may experience an "MFC exception" error during normal operation. If this error occurs, the system will exit the Access 2 user interface (UI) screen and display a window that notifies the user of a "0==I tubes can count" event. If the UI is re-launched, the system will prevent access to the sample manager screen to request samples for processing and the same "MFC exception" error will occur.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime