B) Access 2 Immunoassay Systems- Analyzer Class III
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
81600N
More than 10 numbers, contact manufacturer.
C) Access 2 Immunoassay Systems- Analyzer Class II
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
81600N
800074
A) UniCel DxC 600I Synchron Access Clinical Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
4767
Introduction
Recalled Products
A) UniCel DxC 600I Synchron Access Clinical Analyzer
B) Access 2 Immunoassay Systems- Analyzer Class III
C) Access 2 Immunoassay Systems- Analyzer Class II
D) Access 2 Immunoassay Systems
Reason
Beckman coulter has determined that the Access 2 console PC (B23083) may experience an "MFC exception" error during normal operation. If this error occurs, the system will exit the Access 2 user interface (UI) screen and display a window that notifies the user of a "0==I tubes can count" event. If the UI is re-launched, the system will prevent access to the sample manager screen to request samples for processing and the same "MFC exception" error will occur.