Health Products

OR Table ORT300 (2014-09-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Part #  114093-000OR Table  ORT300
Manufacturer
Imris Inc.
100-1370 Sony PLace
Winnipeg
R3T 1N5
Manitoba
CANADA
114093-000

Introduction

Recalled Products

OR Table  ORT300

Reason

One of the 3 pins to hold an alignment block to the floor of the MR suite was missing until during testing on the weekend the missing pin was pulled into the magnet, damaging the image quality phantom and headrest extension.