Health Products

2008T Hemodialysis System CDX (2012-04-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1TOS108806
Manufacturer
Fresenius USA Inc.
4040 Nelson Avenue
Concord
94520
California
UNITED STATES
2008T Hemodialysis System CDX190713

Introduction

Recalled Products

2008T Hemodialysis System CDX 

Reason

The correction of 10 machines at a customer site was required because they were installed with an unapproved version of software (v 2.30) preloaded on the function board. The version that was installed is approved in the United States but has not yet been reviewed and approved by Health Canada. The newest version is very similar to the currently approved version (v2.13) but does include a change to the user interface and as such requires a Class 3 amendment to the licence before sale in Canada. Due to the similarity between the two versions, the error was not immediately noticeable to the user or patients and thus not observed when the first machines were installed and put in use on Wednesday April 11, 2012.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime