The combination of the test tube heater (K-FTH-1012) and power supply (K-APS-300) was recently re-tested for a proposed design change and found to exceed the limits for radiated emissions, and is therefore no longer compliant with IEC60601-1-2: 2007 (Medical Electrical Equipment - part 1-2 Basic Safety and Essential Performance - Collateral Standard). The radiated RF emissions of the K-APS-300 power supply exceed the limits given in CISPR11 Group 1 Class B by up to 6.3 dBuV/m at three metres with very broad peaks occurring at about 85-90MHz and 150MHz.