Health Products

Digital Diagnost (2014-09-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY
Digital Diagnost9890 010 8390X

Introduction

Recalled Products

Digital Diagnost

Reason

Philips has become aware that in the M-cabinet Componentized X-ray Architecture (CXA) wall box, the protective earth wires to the enclosure are not dimensioned according to Canadian requirements CAN/CSA-C22.2 NO.0.4 "Bonding of Electrical Equipment Impedance Test".