Health Products

Steris 5085 and 5085SRT Surgical Table (2014-10-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0403809001 to 0425514019
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES
B) Steris 5085SRT Surgical TableSTO1-420-X
0403809001 to 0425514019
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES
A) Steris 5085 Surgical TableST01-410-X

Introduction

Recalled Products

A) Steris 5085 Surgical Table
B) Steris 5085SRT Surgical Table

Reason

Steris has learned through customer feedback and field service experience that the shroud cover for the hydraulic column may become misaligned and damaged as a result of extreme table articulation or excessive pressure on the side or base of the table. In most instances, the shroud damage is minor and does not affect the vertical up/down movement of the hydraulic column. In cases of severe shroud damage, the shroud sections may become misaligned such that vertical movement of the hydraulic column is impeded.