Health Products

Bone Level 4.1mm RC, Slactive 8mm TIZR, Loxim Implant (2014-09-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
GY256Bone Level 4.1mm RC, Slactive 8mm TIZR, Loxim Implant
Manufacturer
Institut Straumann AG,
Peter Merian-Weg 12,
Basel
4002
SWITZERLAND
021.4308

Introduction

Recalled Products

Bone Level 4.1mm RC, Slactive 8mm TIZR, Loxim Implant

Reason

It was determined that lot GY256 incorporates the NC version of the Loxim transfer piece rather than the proper RC version. This error could result in the inability to drive the implant into the prepared osteotomy. The transfer piece would turn within the implant and the implant would not advance into the bone.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime