A) Coulter HMX Hematology Analyzers
B) Coulter IH 500 Series System
Reason
Customers have observed that the differential results for coulter 5c abnormal ii cell control recovered an excessive number of incomplete computation (.....) And/or review ("R") flags on the coulter IH 500 series and HMX analyzers. The issue may occur more often when laboratory temperatures fall below 70 degrees Fahrenheit (21 degrees Celsius).