Health Products

Coulter HMX Hematology Analyzers and Coulter IH 500 Series System (2014-10-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B) Coulter IH 500 Series System178833
More than 10 numbers, contact manufacturer.
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A) Coulter HMX Hematology Analyzers6605523

Introduction

Recalled Products

A) Coulter HMX Hematology Analyzers
B) Coulter IH 500 Series System

Reason

Customers have observed that the differential results for coulter 5c abnormal ii cell control recovered an excessive number of incomplete computation (.....) And/or review ("R") flags on the coulter IH 500 series and HMX analyzers. The issue may occur more often when laboratory temperatures fall below 70 degrees Fahrenheit (21 degrees Celsius).