Compass System-Main Software Application (2014-09-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
VERSION COMPASS 3.1
Manufacturer
IBA Dosimetry GmbH
Bahnhofstrasse 5
Schwarzenbruck
90592
GERMANY
Compass System-Main Software ApplicationCS10-100

Introduction

Recalled Products

Compass System-Main Software Application

Reason

The reconstruction of the dose distribution in Compass 3.1 for Monaco-DMLC and Eclipse IMRT plans with jaw tracking enabled result in the same dose distribution as the dose computation, which has the potential of being erroneous. The reason for the wrong result is an error in the software that causes the reconstruction to be based only on the calculated fluence and not on the fluence derived from the measurement.