Health Products

Various Anspach devices (2014-05-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
S) Recipe  M BladeR-SAW
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
H) X max - MotorXMAX
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
I) X max - Foot ControlAUTOLUBE-III
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
F) Sagital Micro BladeS-SAW
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
O) E max 2 Plus System - Console with IrrigationSC2101
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
G) E max - Foot ControlE-FP
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
P) E max 2 Plus System - Hand ControlEMAX2-HC
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
Q) E max 2 Plus System - Foot ControlEMAX2-FP
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
D) Micromax Microsaw AttachmentsMEDIUM-HD
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
E) Adult Rotating CraniotomeCRANI-A-R
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
R) OSC Micro BladeO-SAW
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
B) Micromax AttachmentsADG
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
K) E max Sagittal Saw - MotorMM-SAGITTAL
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
L) Black Max High Performance Instrument System-MotorsBLACKMAX-N
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
C) Micromax Microsaw Keyless DriverDRIVER
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
M) Black Max High Performance Instrument System -Straight AttachmentsB-SILVER
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
N) Black Max High Performance Instrument System -CraniotomeB-GOLD
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
A) Craniotome 5CRANI-A
More than 10 numbers, contact manufacturer.
Manufacturer
The Anspach Effort Inc.
4500 Riverside Drive
Palm Beach Gardens
33410
Florida
UNITED STATES
J) Compact Speed ReducerCSR60

Introduction

Recalled Products

A) Craniotome 5
B) Micromax Attachments
C) Micromax Microsaw Keyless Driver
D) Micromax Microsaw Attachments
E) Adult Rotating Craniotome
F) Sagital Micro Blade
G) E max - Foot Control
H) X max - Motor
I) X max - Foot Control
J) Compact Speed Reducer
K) E max Sagittal Saw - Motor
L) Black Max High Performance Instrument System-Motors
M) Black Max High Performance Instrument System -Straight Attachments
N) Black Max High Performance Instrument System -Craniotome
O) E max 2 Plus System - Console with Irrigation
P) E max 2 Plus System - Hand Control
Q) E max 2 Plus System - Foot Control
R) OSC Micro Blade
S) Recipe  M Blade

Reason

Anspach identified through internal sterility assurance level (SAL) validation testing that several attachments failed to meet 10-6 SAL when sterilized by the Sterrad 100S sterilization method that is specified in the current directions for use (DFU) and operating manuals provided with the device.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime