Health Products

Mammomat Inspiration System - Main Unit (2014-10-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
6007
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Mammomat Inspiration System - Main Unit 101 40 000

Introduction

Recalled Products   

Mammomat Inspiration System - Main Unit 

Reason

Siemens is notifying customers of a potential issue and risk of injury when using the Mammomat inspiration acquisition workstation table when it is not fixed to the floor. If the monitors are positioned too far from the table top, the acquisition workstation table can become unstable with a risk of it falling over and causing injury to the user.