Paradigm Sure-T, Silhouette, Comfort Short, Comfort Kit, Contact Detach, Accu-Check Tender I and II Infusion Sets (2014-11-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.   H) Accu-check Tender II Infusion Set
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
04541464001
More than 10 numbers, contact manufacturer.   F) Contact Detach Infusion Sets
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
10090500
More than 10 numbers, contact manufacturer.   G) Accu-Check Tender I Infusion Set
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
04541413001
More than 10 numbers, contact manufacturer.   D) Comfort Short Infusion Sets
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
10024001
More than 10 numbers, contact manufacturer.  E) Comfort Kit Infusion Sets
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
10000600
All lots,contact manufacturer.B) Silhouette Infusion Sets
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
MMT-371
All lots,contact manufacturer.C) Paradigm Silhouette Infusion  Sets
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
MMT-368600
All lots,contact manufacturer.A) Paradigm Sure-T Infusion Set
Manufacturer
Unomedical A/S
Aaholmvej 1-3,
Osted, Lejre,
4320
DENMARK
MMT-862

Introduction

Recalled Products

   A) Paradigm Sure-T Infusion Set 
   B) Silhouette Infusion Sets
   C) Paradigm Silhouette Infusion Sets 
   D) Comfort Short Infusion Sets
   E) Comfort Kit Infusion Sets
   F) Contact Detach Infusion Sets
   G) Accu-Check Tender I Infusion Set
   H) Accu-Check Tender II Infusion Set

Reason

Manufacturer has identified an increase in reports of the tubing becoming detached at the connect/disconnect location. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify patients. The interruption of insulin delivery can cause hyperglycemia, which if left untreated, can result in diabetic ketoacidosis (DKA). DKA is a serious condition that can cause a severe impact to health, including death. Symptoms of DKA may include nausea, vomiting, shortness of breath and excess thirst/urination. Manufacturer is advising patients to seek medical attention immediately if they are experiencing any of these symptoms.