neoBLUE LED Phototherapy (neoBLUE 2 with replacement board) (2014-11-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Serial# 10507 to 17919A. neoBLUE LED Phototherapy (neoBLUE 2 with replacement board)
Manufacturer
Natus Medical Incorporated,
5900 First Avenue South,
Seattle
98108
Washington
UNITED STATES
010066

Introduction

Recalled products
  1. neoBLUE LED Phototherapy (neoBLUE 2 with replacement board)

Reason

Due to increased efficiency of LED manufacture, the replacement LED boards (part number 001840) for the neoBLUE 2 systems contain LEDs which have higher intensity than those in the original device. As a result, the light intensity may need to be adjusted using the device potentiometer or through treatment distance.  Natus is sending a letter and a copy of the latest technical bulletin to all customers who have purchased a replacement LED board (P/N 001840). Natus will be calling each customer that received the new board to verify the receipt of this letter.