CT TAVI Planning Application in Intellispace Portal V.6 (2014-11-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
550012
Manufacturer
Philips Medical Systems Nederland B.V
Veenpluis 4-6,
Best
5684PC
NETHERLANDS
CT TAVI Planning Application in Intellispace Portal V.6 4598 003 56951

Introduction

CT TAVI Planning Application in Intellispace Portal V.6

Reason

CT Transcatheter Aortic Valve Implantation (CT TAVI) planning application, supplied as an option to Intellispace Portal Software Version 6, may display incorrect measurements of cardiac and aortic anatomy. Clinicians use this information to plan TAVI procedures and to select and size the Transcatheter Heart Valve (THV) to be used in the procedures. Reliance on incorrect information could contribute to an unsuccessful THV implantation procedure.