ABX Pentra 400 Clinical Chemistry Systems (2014-09-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1300602
Manufacturer
Horiba ABX, Parc Euromedecine
rue Du Caducee - BP 7290, Cedex 4,
Montpellier
34184
FRANCE
B) ABX Pentra 400 Clinical Chemistry System ControlsA11A01653
1204103
Manufacturer
Horiba ABX, Parc Euromedecine
rue Du Caducee - BP 7290, Cedex 4,
Montpellier
34184
FRANCE
A) ABX Pentra 400 Clinical Chemistry System - MultiCalA11A01652

Introduction

Recalled Products

A) ABX Pentra 400 Clinical Chemistry System - MultiCal

B) ABX Pentra 400 Clinical Chemistry System Controls

Reason

This recall has been initiated due to the potential that, it was observed that creatinine rate blank method calibrator and control results were reporting values lower than target values indicated in the associated ABX Pentra Multical, N control and P control annexes. Horiba has reassigned the target values and updated annexes for the products and lot numbers.