B) ABX Pentra 400 Clinical Chemistry System Controls
A11A01653
1204103
Manufacturer
Horiba ABX, Parc Euromedecine
rue Du Caducee - BP 7290, Cedex 4,
Montpellier
34184
FRANCE
A) ABX Pentra 400 Clinical Chemistry System - MultiCal
A11A01652
Introduction
Recalled Products
A) ABX Pentra 400 Clinical Chemistry System - MultiCal
B) ABX Pentra 400 Clinical Chemistry System Controls
Reason
This recall has been initiated due to the potential that, it was observed that creatinine rate blank method calibrator and control results were reporting values lower than target values indicated in the associated ABX Pentra Multical, N control and P control annexes. Horiba has reassigned the target values and updated annexes for the products and lot numbers.