Health Products

Dialog+ SW 9.XX Two Pump Machines (incl. HDF online) (2014-11-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.A) Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)
Manufacturer
B. Braun Avitum AG
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY
710201CV

Introduction

Recalled Products

Dialog+ SW 9.XX Two Pump Machines (incl. HDF online)

Reason

There is a risk of air infusion into the arterial line if the valve block level regulation and/or diaphragm pump is installed incorrectly during technical service. The correct functionality of manual regulation must be tested after service. The notification affects only units in which the valve block has been exchanged.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime