Dimension Vista System - MYO Calibrator (2014-11-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4FD085
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
Dimension Vista System - MYO CalibratorKC624

Introduction

Recalled Products

Dimension Vista System - MYO Calibrator

Reason

The Dimension Vista Myoglobin Calibrator (MYO CAL) lot 4FD085 may produce a positive shift in MYO QC and patient test results that exceeds their acceptance criteria for this product. Siemens has observed a positive shift up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.