Medionics has received a few reports of opacification and ballooning of the peritoneal dialysis catheters at the exit-site, however no adverse events have been reported associated with this issue. The age of the implanted catheters in the reported incidents ranged from 30 months to 80 months.
Investigation of these incidents indicates that the structural changes to PD catheters at the exit-site may occur with prolonged use of certain cleansing agents and antibiotic cream or ointment. Additionally, use of solutions containing organic solvents for catheter cleaning may cause the breakdown of silicone tubing. The likelihood of structural changes to silicone PD catheters increases over the duration of implantation.