Vitros Immunodiagnostic, Integrated and Chemistry System (2014-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
36000130
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
D) Vitros 3600 Immunodiagnostic System (Licence # 80702)680 2783
More than 10 numbers, contact manufacturer.  
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
E) Vitros 4600 Chemistry System680 2445
More than 10 numbers, contact manufacturer.  
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B) Vitros 5600 Integrated System (Licence # 78571)680 2413
More than 10 numbers, contact manufacturer.   
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
C) Vitros 5600 Integrated System (Licence # 80701)680 2413
36000130
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A) Vitros 3600 Immunodiagnostic System (Licence # 78572)680 2783

Introduction

Recalled Products

A) Vitros 3600 Immunodiagnostic System (Licence # 78572)
B) Vitros 5600 Integrated System (Licence # 78571)
C) Vitros 5600 Integrated System (Licence # 80701)
D) Vitros 3600 Immunodiagnostic System (Licence # 80702)
E) Vitros 4600 Chemistry System

Reason

Ortho Clinical Diagnostics (OCD) has identified an anomaly with Vitros system software version 3.1 and below. OCD determined that the software may not properly identify an expired calibration. Calibration status is monitored by the system software which is based upon the calibration interval and the date of the last calibration. OCD's investigation confirmed that if the anomaly occurs, a calibration may expire, but the system may not alert the operator and not flag results with a CE sample result code (i.e. Calibration expired). The system may unexpectedly generate results even though it is configured to not report results when a calibration has expired.