Health Products

ADVIA 1200 Chemistry System - Liquid Specific Protein Calibrator and ADVIA Chemistry System - Liquid Specific Protein (2014-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
303574
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B) ADVIA Chemistry System - Liquid Specific Protein07711199
303574
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A) ADVIA 1200 Chemistry System - Liquid Specific Protein Calibrator07711199

Introduction

Recalled Products

A) ADVIA 1200 Chemistry System - Liquid Specific Protein Calibrator
B) ADVIA Chemistry System - Liquid Specific Protein

Reason

Siemens Health Care Diagnostics has confirmed two issues with the Siemens liquid specific protein calibrator lot 303574.

Issue 1: increased frequency of max fit deviation flags using the liquid specific protein (LSP) calibrator kit lot 303574 (vial lots 122IT-127IT) to calibrate the alpha-1-antitrypsin (ATT) assay. The max fit deviation flag impacts the level 2 calibrator (123IT) for the Advia Chemistry Systems AAT assay.

Issue 2: Siemens confirmed that the preliminary version of the Siemens LSP calibrator lot-specific instructions for use (IFU) for calibrator kit lot 303574 resided in the Siemens document library from Dec.10, 2013 until Apr.07, 2014. The LSP calibrator preliminary factor values during this time differed slightly from the final factor values for the following Advia Chemistry System assays: AAG, AAT, ASO_2, C3, C4, FRT, HAPT, IgA_2, IgG_2, IgM_2, PreALB, RF and TRF. The IFU insert packaged in the Siemens LSP calibrator kit lot 303574, always contained the correct values for all those assays, except AAT. The IFU that is currently in the kit and the IFU that was in the Siemens document library from Apr.07, 2014 to Oct.30, 2014 does not reflect the updated AAT calibrator value for LSP calibrator level 2. As of Oct.31, 2014 all of the final LSP calibrator values are correct on the Siemens document library. When using either the final or preliminary calibrator values, quality control results recover well within the manufacturer's specified upper and lower ranges and patient sample results did not demonstrate clinically significant differences.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime