Heartstart XL,MRx,XL+ Defibrillator/Monitor and ECG Out Sync Cable (2014-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
D) ECG Out Sync CableM1783A
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
B) Heartstart MRx Defibrillator/MonitorM3535A
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
C) Heartstart XL+861290
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
A) Heartstart XLM4735A

Introduction

Recalled Products

A) Heartstart XL
B) Heartstart MRx Defibrillator/Monitor
C) Heartstart XL+
D) ECG Out Sync Cable

Reason

Philips has become aware that when a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTS) while connected to ac power:
  • On the Heartstart MRx and Heartstart XL, EFT noise can be misinterpreted as a R-wave
  • on the Heartstart XL+, EFT noise can disable ECG monitoring, and potentially interrupt demand mode pacing*.

1. *Note: it is contrary to the instructions in the XL+ instructions for use (IFU) to perform demand mode pacing while using the sync cable to provide the ECG signal from a beside monitor. The XL+ IFU includes the following warning: ""when pacing in demand mode, the ECG cable from the patient must be directly connected to the Heartstart XL+."" if the user follows this warning, this problem cannot occur on the XL+.