Various Bronchial Double lumen Tube Sets (2014-11-21) Nov 21, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number 13HT12 D) White Bronchial Double Lumen Tube Set (Right) Manufacturer Teleflex Medical IDA Business and Technology Park, Dublin Road, Athlone, CO. Westmeath IRELAND 116201350 More than 10 numbers, contact manufacturer. B) Bronchial Double Lumen Tube Set (Right) Manufacturer Teleflex Medical IDA Business and Technology Park, Dublin Road, Athlone, CO. Westmeath IRELAND 116200260 More than 10 numbers, contact manufacturer. C) Carlens Bronchial Double Lumen Tube Set (Left) Manufacturer Teleflex Medical IDA Business and Technology Park, Dublin Road, Athlone, CO. Westmeath IRELAND 116101370 More than 10 numbers, contact manufacturer. A) Bronchial Double lumen Tube Set (Left) Manufacturer Teleflex Medical IDA Business and Technology Park, Dublin Road, Athlone, CO. Westmeath IRELAND 116100260
Recalled Products
A) Bronchial Double lumen Tube Set (Left)
B) Bronchial Double Lumen Tube Set (Right)
C) Carlens Bronchial Double Lumen Tube Set (Left)
D) White Bronchial Double Lumen Tube Set (Right)
Teleflex medical is recalling the product following receipt of customer complaints reporting that the cobb connector detached from the main connector prior to use.