Galileo NEO Automated Blood Bank System (2014-12-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All serial numbers.Galileo NEO Automated Blood Bank System (2014-12-11)
Manufacturer
Immucor Inc.
3130 Gateway Drive,
0064599

Introduction

Recalled Products

Galileo NEO Automated Blood Bank System

Reason

Following the launch of the last software upgrade to neo install set 2.1 (Aurora 2.09), there have been sporadic complaints in which results are not being received in the data management software (DMS). Users most commonly report that results are not visible in the results folder and/or do not print once processing is complete. When this occurs, the user must intervene to try to resend the results to the DMS or in many cases repeat the test. In this event, the hazard is delay in reporting results. This issue has been linked to reuse of capture plate frames and users are advised to not use a plate frame more than 12 times.