Health Products

Persona TASP Shims and Persona Constrained TASP Shims (2014-12-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
62622533
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
B) Persona Constrained TASP Shims42-5279-002-00
More than 10 numbers, contact manufacturer.
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
A) Persona TASP ShimsMore than 10 numbers, contact manufacturer.

Introduction

Recalled Products   
  1. Persona TASP Shims
  2. Persona Constrained TASP Shims

Reason

Zimmer is initiating this notification due to the discovery of persona Tibial Articular Surface Provisional (TASP) shims, with missing ball bearings. Repeated cycles using sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this from occurring, persona TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.