Health Products

Immage Immunochamistry Systems Apolipoprotein A-1 (APA) and Beckman Coulter Clinical Chemistry Systems APO CAL (Apolipoprotein Calibrator) (2014-12-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M305133
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Beckman Coulter Clinical Chemistry Systems APO CAL (Apolipoprotein Calibrator)469283
M404171
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. Immage Immunochamistry Systems Apolipoprotein A-1 (APA)446410

Introduction

Recalled Products
  1. Immage Immunochamistry Systems Apolipoprotein A-1 (APA)
  2. Beckman Coulter Clinical Chemistry Systems APO CAL (Apolipoprotein Calibrator)

Reason

Beckman coulter has confirmed the Immage Immunochemistry Systems Apolipoprotein A-1 (APA) reagent (lot number M404171) contains an important product notification insert with incomplete information. The notification references only APA lot number M401225, but should reference APA lot numbers M401225 and newer. APA lot number m404171 calibrated with APO CAL lot number M305133 may produce inaccurate results (>6% low bias).