Phillips IntelliVue Information Centre (PIIC) IX (2014-12-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
866023
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
Phillips IntelliVue Information Centre (PIIC) IX M3290B

Introduction

Recalled Products

Phillips IntelliVue Information Centre (PIIC) IX 

Reason

A software defect that could potentially result in a drop in oxygenation (hypoxia) that is not immediately detected by the clinician. Defect involves the use of the "change bed label" feature at the information center ix. If the configuration for this item is set to "optional" and a user selects the "location" button within the "manage patient" menu and makes a change to the patient location, alarms for sp02 and/or non-invasive Blood Pressure (NBP) will become disabled without visual notification (bell with an "x") to the user. This problem only happens when the patient is monitored using a networked Intellivue TRX Telemetry M4841/TRX 4851a patient worn device (PWD).