Heartstart MRx Defibrillator/Monitor (2014-12-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
FOR MODEL M3535A:Heartstart MRx Defibrillator/Monitor 
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
M3535A

Introduction

Recalled Products   

Heartstart MRx Defibrillator/Monitor 

Reason

Philips has become aware that the Heartstart MRx Defibrillator/Monitor could reboot once in any operating mode if the following sequence occurs:
  1. In service mode, the status log* is cleared
  2. Prior to exiting service mode, the device generates an entry in the status log.
  3. Service mode is exited.
The Heartstart MRx device is then in a state that could trigger a reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed. The Heartstart MRx device will restart within 8 seconds. Note: if a reboot occurs during clinical use, current device settings and patient record are retained. *the status log is a tool accessed by service personnel. It includes entries for all messages logged during normal operating mode, automated tests, service and configuration mode, and operational checks. The Heartstart MRx defibrillator/monitor service manual instructs service personnel to clear the status log after successful completion of an operational check.