Health Products

Monaco (2014-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable  
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems Inc.
13723 Riverport Drive, Suite 100
Maryland Heights
63043
Missouri
UNITED STATES
Monaco MONACO

Introduction

Recalled Products   

Monaco 

Reason

In the case where CT slice thickness is not the same as CT slice spacing the image statistics tool will report an incorrect HU. If the user bases their CT to ED file on this incorrect density information, dose calculation errors will result. The magnitude of the dose calculation error will be dependent on the magnitude of the HU error.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime