Health Products

MSK Extreme and Optima MR430S 1.5T MRI Systems (2014-12-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
514697ONI
Manufacturer
GE Medical Systems, LLC
301 Ballardvale St., Suite 4,
B) Optima MR430S 1.5T - MRI SystemOPTIMA MR430S 1.5T (5000-0002)
905521ONI
Manufacturer
GE Medical Systems, LLC
301 Ballardvale St., Suite 4,
A) MSK Extreme - MRI SystemAB2000

Introduction

Recalled Products

A) MSK Extreme - MRI System
B) Optima MR430S 1.5T - MRI System

Reason

Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.