Health Products

Immulite, Immulite1000 and Immulite 2000 Systems (2014-12-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
B) Immulite 1000 System-CEA AssayLKCE1
All lots.
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
C) Immulite 2000 System-CEA AssayL2KCE2
All lots.
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
A) Immulite System-CEA AssayLKCE1

Introduction

Recalled Products

A) Immulite System-CEA Assay
B) Immulite 1000 System-CEA Assay
C) Immulite 2000 System-CEA Assay

Reason

Siemens internal investigation confirmed that the Immulite/Immulite 1000 and the Immulite 2000/ 2000 XPi CEA assay does not meet the (within-run precision specifications as per the Immulite/Immulite 1000 and the Immulite 2000/ 2000 XPi IFU's for all in date Immulite CEA kit lots. The Immulite/Immulite 1000 and the Immulite 2000/ 2000 XPi CEA inter-assay (total precision) shows equivalence to the IFU and is therefore not impacted by this issue.