Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue,
Chatsworth
91311
California
UNITED STATES
B) RT12 Rotor KitNot applicable  
All lots.
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue,
Chatsworth
91311
California
UNITED STATES
A) Statspin PM Centrifuge (SSMP)M901-22

Introduction

Recalled Products

A) Statspin PM Centrifuge (SSMP)
B) RT12 Rotor Kit

Reason

Iris International has received reports of uncontained debris (part fragments) as a result of rotor breakage and lid failures on similar centrifuge models which use the RT12 rotor and shield.

If the RT12 rotor breaks and the centrifuge fails to contain debris, escaping debris may result in possible operator injury, direct exposure to biohazardous materials, and/or loss of sample. Rotor breakage may also result in damage to the centrifuge.