Health Products

Signa HDE 1.5T and Discovery MR750W 3.0T MRI Systems (2014-12-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
403229MR1
Manufacturer
GE Healthcare Japan Corporation
7-127, Asahigaoka, 4-Chome,
Hino-shi, Tokyo
191-8503
JAPAN
B) Discovery MR750W 3.0T MRI SystemDISCOVERY MR750W 3.0T
450688MR
Manufacturer
GE Healthcare Japan Corporation
7-127, Asahigaoka, 4-Chome,
Hino-shi, Tokyo
191-8503
JAPAN
A) Signa HDE 1.5T MRI SystemS7505KY

Introduction

Recalled Products

A) Signa HDE 1.5T MRI System
B) Discovery MR750W 3.0T MRI System

Reason

Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.