Optima MR360 Main Unit (2014-12-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
403541MR
Manufacturer
GE Hangwei Medical Systems Co. Ltd.
No. 2 North Yong Chang Street, Beijing Economic, Technological Development Zone,
Beijing
100176
CHINA
Optima MR360 Main Unit5312366 (M50002BB)

Introduction

Recalled Products

Optima MR360 Main Unit

Reason

Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.