Health Products

AU680 and AU5800 Clinical Chemistry Analyzer (2014-12-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2013020679 to 2014121891
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B) AU5800 Clinical Chemistry AnalyzerAU5810
2013022088 to 2014012894
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A) AU680 Clinical Chemistry AnalyzerAU680

Introduction

Recalled Products

A) AU680 Clinical Chemistry Analyzer
B) AU5800 Clinical Chemistry Analyzer

Reason

Beckman coulter has confirmed the au680 and au5800 clinical chemistry analyzers are not fully compliant with the applicable EMC standard (Electromagnetic Disturbance Characteristics CISPR 11/group 1, class a). The investigation revealed that a previous design change made to the cooling module resulted in the noise level of the new cooling module exceeding the EMC standard limit.