Health Products

Mevatron, Oncor Expression Digital Linear, Artiste MV Linear Accelerators and Oncor Impression Plus (2014-12-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3799
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
D) Oncor Impression Plus5857912
5037
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen, Germany
80333
B) Oncor Expression Digital Linear Accelerator7360717
5414
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen, Germany
80333
C) Artiste MV - Linear Accelerator8139789
2559
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A) Mevatron Linear Accelerators01940035

Introduction

Recalled Products

A) Mevatron Linear Accelerators
B) Oncor Expression Digital Linear Accelerator
C) Artiste MV - Linear Accelerator
D) Oncor Impression Plus

Reason

Siemens has received reports from one electric board (G41) located inside of the linac gantry showing signs of overheating. After a detailed analysis the root cause was found to be one of the components that belong to this board (head driver PCB).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime