Health Products

DiscPac Tip Caps (2014-12-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Baxa Corporation
14445 Grasslands Drive
Englewood
80112
Colorado
UNITED STATES
DiscPac Tip CapsMore than 10 numbers, contact manufacturer.

Introduction

Recalled Products

DiscPac Tip Caps

Reason

During a label review, it was identified that the single and case label templates for DiscPac tip caps did not include the latex symbol. DiscPacs were shipped with a missing latex symbol on both single and case label artwork. The IFU correctly identifies the latex content. The heat shrink tape used on the DiscPac package is the only component that contains latex. The tape does not come into direct contact with the tip caps, but must be peeled away and removed to access the self-righting tip caps.