Health Products

CD8 PE (Leu 2a) Reagent (2014-12-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4220622
Manufacturer
Becton, Dickinson and Company
BD Biosciences
2350 Qume Drive
San Jose
95131
California
UNITED STATES
CD8 PE (Leu 2a) Reagent340046

Introduction

Recalled Products

CD8 PE (Leu 2a) Reagent

Reason

May be hazardous to health: The CD8 PE reagent is contaminated with CD4 FITC. CD4 cells will be expressed along the FITC axis. Any marker that is stained for on the FITC axis will contain this incorrectly stained CD4 population. With subsequent down gating, the results will appear as if the CD4 population is not present. This could result in the unnecessary intensification of therapy for patients already known to have reductions in their CD4 levels.