From lot # 178364 and higher
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
Mannheim
68305
GERMANY
PreciControl Multimaker05341787190

Introduction

Recalled Products

PreciControl Multimaker

Reason

During internal stability testing of the reconstituted PreciControl Multimaker of lot 178364, a reduced stability of PIGF became apparent. Whereas the stability claim for all analytes in PreciControl Multimaker is 72 hours at 2-8c, this claim does not apply for PIGF As a consequence, the stability of PIGF has been reduced accordingly in the method sheet of the PreciControl Multimaker. The claim at 2-8c was removed but the frozen and ambient temperature stability remained the same as they are not affected.