A) LICOX Micro-Probe Kit
B) LICOX Brain Tissue Oxygen Monitoring System - Complete Brain Probe Kit
C) LICOX Brain Tissue Oxygen Monitoring System - Complete Brain Tunneling Probe Kit
D) Combined Oxygen and Temperature Probe
E) Oxygen Probe
Reason
Integra Lifesciences initiated a recall to remove all affected LICOX probes and probe kits that may contain an inaccurate smart cards supplied with the subject devices. The IKN2 value in approximately 90% of the potentially affected cards is approximately 10 times higher than it should be resulting in varying degrees of understated PO2 values (from -1 mm of Hg to -7 mm of Hg) when the probes are used.