Health Products

Esprit Ventilator (2014-12-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Respironics California, Inc.
2271 Cosmos Court
Carlsbad
92011
California
UNITED STATES
Esprit VentilatorV1000

Introduction

Recalled Products

Esprit Ventilator

Reason

Esprit V1000 manufactured prior to December 31, 2004 may have the power supply fan mounting screws installed incorrectly, which could cause the ends of the screws to touch the power supply PCB-A and short to ground.