Health Products

Discovery IGS (2014-01-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
AM12P3D241
Manufacturer
GE Medical Systems SCS
283 Rue De La Miniere
Buc
78530
FRANCE
Discovery IGSS18631BA (5418812)

Introduction

Recalled Products

Discovery IGS

Reason

Systems may experience an unintentional motion or rotation due to a positioning user interface issue. The potential reversed motion of the gantry/table has been observed when the user positions the joystick at its maximum position (end of travel) on the smart box. The unintended reversal of the direction of the motion may lead to a collision of the table or C-arm with medical instruments, patient or personnel, and result in injury.