Health Products

ABL90 Flex - Analyzer (2014-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All serial numbers.       ABL90 Flex – Analyzer 
Manufacturer
Radiometer Medical APS
Akandevej 21
Bronshoj
2700
DENMARK
393-090

Introduction

Recalled Products

ABL90 Flex – Analyzer

Reason

Radiometer has recently become aware of a potential significant clinical issue with the results reported if or when the inlet is left open after aspiration of the patient sample, for more than 30 seconds.