Health Products

Statim 7000 Cassette Sterilizer (2015-01-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
SciCan Ltd.
1440 Don Mills Road
Toronto
M3B 3P9
Ontario
CANADA
A. Statim 7000 Cassette Sterilizer01-702100

Introduction

Recalled products
  1.   Statim 7000 Cassette Sterilizer

Reason

This recall has been initiated due to a part that is no longer available. This part is an RFID (radio-frequency identification) tag which is used in the chamber seal to monitor the number of cycles ran in order to inform the user when it is necessary to change the chamber seal. Our current supplier of this part has decided to exit the RFID tag business and a suitable replacement was not found. Therefore, all future cassette seals will no longer contain the RFID tag (seals are regularly replaced by the user). When this tag is not present, the machine will not allow a cycle to start with the current software. LN order to implement this change in the Statim 7000 and remove the RFID tag from the cassette seal, the RFID board software must be updated to allow the unit to run without an RFID tag in the cassette seal. This will be performed by trained technicians at the user site. There is no need for the unit to be returned to the manufacturer for this software update. The life of the seal is still 1000 cycles or every twelve months, whichever comes first, but no notification will be given on the unit once the tag is no longer present and the software has been updated. It will be the user's responsibility to monitor the number of cycles ran and the seal's performance. Manual monitoring of seal replacement is a common practice for most steam sterilizers on the market.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime